Determination of CD4 and CD8 lymphocyte subsets by a new alternative fluorescence immunoassay

T. N. Denny, B. D. Jensen, E. I. Gavin, A. G. Louzao, F. A. Vella, J. M. Oleske, W. Wong

Research output: Contribution to journalArticle

9 Scopus citations

Abstract

The purpose of this study was to evaluate a new alternative fluorescence immunoassay method (Zymmune CD4/CD8 Cell Monitoring Kit; Zynaxis, Inc., Malvern, Pa.) for determining the absolute CD4+ and CD8+ T-lymphocyte concentrations in whole blood. The investigation was performed as a two-site comparison of the reference whole blood flow cytometric method with the Zymmune method. In this investigation, a total of 166 patient samples were evaluated of which ~ 20% were from human immunodeficiency virus-positive individuals. The mean value for samples performed by the Zymmune CD4 assay was 1,094 (range, 74 to 2,586) cells per μl, while the reference method yielded a mean of 890 (range, 35 to 2,033) cells per μl. The correlation coefficient for regression analysis was 0.940. The mean value for samples performed by the Zymmune CD8 assay was 700 (range, 212 to 1,813) cells per μl, while the reference method yielded a mean of 546 (range, 82 to 2,158) cells per μl. The correlation coefficient for regression analysis was 0.921. No site-specific differences or trends in CD4 or CD8 values were seen when the data were analyzed by site of collection. The average precision of the CD4 assay varied from 6 to 14%, corresponding to the high and low concentration ranges. For CD8, the average precision varied from 8.3 to 16% over the respective high to low concentration ranges. We conclude that the Zymmune CD4/CD8 Cell Monitoring Kit method provides absolute CD4+ and CD8+ T-lymphocyte concentrations which are equivalent to those given by the reference flow cytometric method.

Original languageEnglish (US)
Pages (from-to)330-336
Number of pages7
JournalClinical and Diagnostic Laboratory Immunology
Volume2
Issue number3
DOIs
StatePublished - 1995

All Science Journal Classification (ASJC) codes

  • Microbiology (medical)
  • Clinical Biochemistry
  • Immunology and Allergy
  • Immunology

Fingerprint Dive into the research topics of 'Determination of CD4 and CD8 lymphocyte subsets by a new alternative fluorescence immunoassay'. Together they form a unique fingerprint.

  • Cite this